Basic HTML Version

sec
SEC
33
SUMMARY OF TSKgel SIZE EXCLUSION COLUMN LINES
Column line
TSKgel SW / SW
XL
/
SuperSW
TSKgel PW / PW
XL
TSKgel Alpha /
TSKgel SuperAW
TSKgel H
Particle composition
Silica
Polymethacrylate
highly crosslinked Poly-
methacrylate
PS-DVB
No. of available pore
sizes
3/2
7
5
6
pH stability
2.5 - 7.5
2.0 - 12.0
2.0 - 12.0
1.0 - 14.0
Solvent compatibility
100% polar
50% polar
100% polar and
nonpolar
100% nonpolar, limited
polar
Max. temperature
30°C
80°C*
80°C
60-80°C (H
XL
and
SuperHZ)
140°C (H
HR
and SuperH)
Max flow rate
(mL/min)
6.0 (SW, SW
XL
) /
0.4 (SuperSW)
1.2 (PW) / 1.0 (PW
XL
)
1.0 (Alpha) / 0.6 (SuperAW)
Pressure** (MPa)
1.0-12.0
1.0 - 4.0
2.0 - 4.0
15-60
Application focus
proteins
water-soluble polymers
intermediate polar
polymers
organic-soluble
polymers
*
Except for the TSKgel G-DNA-PW, which can be operated up to 50°C and the 55 mm ID TSKgel PW-type columns, which can be
operated up to 60°C. When operating below 10°C, reduce the flow rate to ensure that the maximum pressure is not exceeded.
**
Depends on column dimensions and particle size.
Note: The operating conditions and specifications for each column are listed on the Operating Conditions and Specifications sheet (OCS)
shipped with the column and in the Ordering Information section at the end of each section.