Size Exclusion: Gel Permeation Chromatography
TSK-GEL Columns
Proteins (biopolymers)
Organic-soluble Polymers
TSKgel Hxl
TSKgel Hhr
TSKgel SuperH
TSKgel SuperHZ
TSKgel SuperMultipore
Water-soluble Polymers
Polar Organic-Soluble Polymers


TSKgel GPC columns
are available in conventional 7.8 mm ID x 30 cm L Hxl and Hhr columns and in high efficiency, high throughput 6 mm ID x 15 cm L SuperH and SuperHZ columns. How the product lines compare to one another is shown below.

   Comparison of TSKgel H-type resins

Application focus

High-throughput polymer analysis with ultra low polymer adsorption, limited solvent compatibility range.

Conventional polymer analysis with ultra low polymer adsorption, limited solvent compatibility range

High-throughput polymer analysis with expanded solvent compatibility range.

Conventional polymer analysis with expanded solvent compatibility range.

Particle size

3, 5, and 10 micron, depending on pore size

5 and 9 micron, depending on pore size

3 and 5 micron, depending on pore size

5 micron

Theoretical plates 1

16,000/15 cm column

16,000/30 cm column

16,000/15 cm column

16,000/30 cm column

Maximum temperature

G1000--G3000, 60°C
G4000--Mixed, 80°C

G1000--G3000, 60°C
G4000--Mixed, 80°C

140°C

140°C

Standard shipping solvent

THF

THF2

THF

THF2

THF can be switched to

see table 1 (below) and note #3 for dichloromethane and dichloroethane

see table 1 (below)

Other shipping solvents available?

yes4

no

Number of solvent substitutions

One time only

Several5

Solvent exchange instructions

Linear gradient with a 2%/min rate of change with a flow rate of <0.25 mL/min.

 

Linear gradient with a 2%/min rate of change with a flow rate of <0.5 mL/min.

Linear gradient with a 2%/min rate of change according to flow rates in table 2 (below).

1) Theoretical plates listed are based on smallest particle size listed.
2) High-temperature columns (HT) are shipped with ODCB (o-dichlorobenzene) as standard shipping solvent.
3) Dichloromethane and dichloroethane are not available for G1000 pore size columns.
4) See Table 1 (below) for available shipping solvents
5) After switching to a very polar solvent such as acetone, switching back to a nonpolar solvent is not recommended.

   Table 1: Solvent compatibility for TSKgel H-type columns
The column shipping solvent limits the types of solvents that can be used with each column type.

TSKgel SuperHZ and Hxl1

Tetrahydrofuran3

benzene, chloroform, toluene, xylene, dichloromethane, dichloroethane

Acetone**

carbon tetrachloride5, o-dichlorobenzene, dimethylformamide, dodecane, dimethyl sulfoxide, dioxane, ethylacetate, FC-113, hexane, pyridine, hexafluoroisopropanol/chloroform, methyl ethyl ketone, quinoline, cyclohexane

Chloroform**

m-cresol in chloroform, up to 10% hexafluoroisopropanol/chloroform

Dimethylformamide dimethyl sulfoxide, dioxane, tetrahydrofuran, toluene
o-Dichlorobenzene 1-chloronaphthalene, trichlorobenzene
TSKgel SuperH and Hhr2 Tetrahydrofuran3

acetone, ethanol, quinoline, benzene, o-dichlorobenzene, ethyl acetate, dodecane, FC-113, carbon tetrachloride5, dichloromethane, dichloroethane, trichloroethane, n-hexane, cyclohexane, xylene, tetrahydrofuran, chloroform, 1,4-dioxane, hexafluoroisopropanol, toluene, 1-chloronaphthalene, dimethylaceto acetamide, methyl ethyl ketone, trichlorobenzene, m-cresol, dimethylformamide, methylpyrrolidone, o-chlorophenol/chloroform, dimethyl sulfoxide, pyridine

 

Notes:

1) In case of TSKgel SuperHZ and Hxl, keep flow rate below <0.5 mL/min during solvent change. Solvent can be changed one way/one time only.
2) In case of TSKgel SuperH and Hhr, see Table 2 below for appropriate flow rates for solvent exchange. After switching to a very polar solvent, switching to a nonpolar solvent is not recommended.
3) All TSKgel Hxl, Hhr, SuperHZ, SuperH and GMH analytical columns are shipped containing tetrahydrofuran (THF), except TSKgel GMH-HT columns, which contain o-dichlorobenzene (ODCB).
4) THF in TSKgel G1000Hxl columns cannot be replaced with dichloromethane or dichloroethane.
5) Prolonged exposure to carbon tetrachloride can corrode the stainless steel parts of a column and an HPLC system.

*  100% methanol cannot be used with H-type columns; use this solvent with TSKgel SW-type or Alpha-type columns.

** TSKgel H columns may be specially ordered with this shipping solvent.

   Table 2: Recommended flow rates (mL/min) for solvent exchange in TSKgel SuperH and Hhr columns
Solvent SuperH (6.0mmIDx15cm) Hhr (7.8mmIDx30cm)

n-Hexane

0.5

0.9

methyl ethyl ketone

0.4

0.7

dichloromethane, ethyl acetate

0.35

0.6

toluene, chloroform

0.3

0.5

dimethylformamide

0.2

0.4

carbon tetrachloride, pyridine

0.15

0.3

dimethyl sulfoxide, dioxane, ethanol,
N-methylpyrrolidone, o-dichlorobenzene

0.1

0.2

quinoline, hexafluoroisopropanol,
1-chloronaphthalene

0.05

0.1


Comparing SuperHZ with Hhr Column

Size Exclusion main page

Hxl Main page

Hhr main page

SuperH main page

SuperHZ main page